The Reduction of fumecupboard running costs
Contenido multimedia no disponible por derechos de autor o por acceso restringido. Contacte con la institución para más información.
Tag | 1 | 2 | Value |
---|---|---|---|
LDR | 00000nab a2200000 i 4500 | ||
001 | MAP20071017413 | ||
003 | MAP | ||
005 | 20080418115041.0 | ||
007 | hzruuu---uuuu | ||
008 | 940216s1982 gbr|||| | |00010|eng d | ||
035 | $a6800003269 | ||
040 | $aMAP$bspa | ||
084 | $a870 | ||
100 | 1 | $0MAPA20080012496$aHughes, D. | |
245 | 1 | 4 | $aThe Reduction of fumecupboard running costs$cD. Hughes and B. McIntosh |
520 | $aThe power required to move and to heat the air consumed by a laboratory fumecupboard is considered. A significant economy is obtained if the volume rate of air extracted can be halved. Such a reduction is acceptable if it occurs when the sash is closed to half the maximum aperture so that the design face velocity is maintained. This can achieved by coupling the fan to a two-speed electric motor controlled by a limit switch automatically actuated by the fumecupboard sash | ||
650 | 1 | 1 | $0MAPA20080585679$aHigiene industrial |
650 | 1 | 1 | $0MAPA20080604967$aCampanas de extracción |
650 | 1 | 1 | $0MAPA20080605278$aContaminantes químicos |
650 | 1 | 1 | $0MAPA20080565954$aHumos tóxicos |
650 | 1 | 1 | $0MAPA20080612597$aSistemas de ventilación |
650 | 1 | 1 | $0MAPA20080560997$aVentilación |
650 | 1 | 1 | $0MAPA20080579784$aCostes económicos |
700 | 1 | $0MAPA20080076603$aMacIntosh, B. | |
740 | 4 | $aThe Annals of occupational hygiene | |
773 | 0 | $tThe Annals of occupational hygiene$dOxford [etc.]$gnº 1, 1982 ; p. 101-104 |