Search

A New method for quantitative : in-use testing of laboratory fume hood

Fichero PDF / PDF file
MARC record
Tag12Value
LDR  00000nab a2200000 i 4500
001  MAP20071019428
003  MAP
005  20080418115812.0
007  hzruuu---uuuu
008  940601e19890501usa|||| | |00010|eng d
035  ‎$a‎6800013111
040  ‎$a‎MAP‎$b‎spa
084  ‎$a‎872
1001 ‎$0‎MAPA20080195373‎$a‎Ivany, Raymond E.
24512‎$a‎A New method for quantitative‎$b‎: in-use testing of laboratory fume hood‎$c‎Raymond E. Ivany, Melvin W. First, and Louis J. Diberardinis
520  ‎$a‎A new method has been developed for testing fume hood performance in a controlled, ventilation laboratory setting or in the field during normal use conditions. It involves a steady release of a tracer material, such a sulfur hexafluoride, inside a hood through a hollow rectangular diffuser that surrounds the work and then measurement of concentration outside the hood
65011‎$0‎MAPA20080585679‎$a‎Higiene industrial
65011‎$0‎MAPA20080616441‎$a‎Contaminación atmosférica
65011‎$0‎MAPA20080605278‎$a‎Contaminantes químicos
65011‎$0‎MAPA20080606411‎$a‎Hexafluoruro de azufre
65011‎$0‎MAPA20080621186‎$a‎Control de la contaminación
65011‎$0‎MAPA20080586355‎$a‎Métodos de control
65011‎$0‎MAPA20080543815‎$a‎Ensayos
65011‎$0‎MAPA20080563134‎$a‎Laboratorios
65011‎$0‎MAPA20080604967‎$a‎Campanas de extracción
7001 ‎$0‎MAPA20080162429‎$a‎First, Melvin W.
7001 ‎$0‎MAPA20080299583‎$a‎Diberardinis, Louis J.
7400 ‎$a‎American Industrial Hygiene Association journal
7730 ‎$t‎American Industrial Hygiene Association journal‎$d‎Akron, Ohio‎$g‎Vol. 50, nº 5, May 1989 ; p. 275-280