Búsqueda

The Reduction of fumecupboard running costs

Fichero PDF / PDF file
MAP20071017413
Hughes, D.
The Reduction of fumecupboard running costs / D. Hughes and B. McIntosh
Sumario: The power required to move and to heat the air consumed by a laboratory fumecupboard is considered. A significant economy is obtained if the volume rate of air extracted can be halved. Such a reduction is acceptable if it occurs when the sash is closed to half the maximum aperture so that the design face velocity is maintained. This can achieved by coupling the fan to a two-speed electric motor controlled by a limit switch automatically actuated by the fumecupboard sash
En: The Annals of occupational hygiene. - Oxford [etc.]. - nº 1, 1982 ; p. 101-104
1. Higiene industrial . 2. Campanas de extracción . 3. Contaminantes químicos . 4. Humos tóxicos . 5. Sistemas de ventilación . 6. Ventilación . 7. Costes económicos . I. MacIntosh, B. . II. Título. III. Título: The Annals of occupational hygiene.