Biological safety cabinets : decontamination or sterilization with paraformaldehyde

Fichero PDF / PDF file
Registro MARC
Tag12Valor
LDR  00000nab a2200000 i 4500
001  MAP20071020019
003  MAP
005  20080418120044.0
007  hzruuu---uuuu
008  940623e19880601usa|||| | |00010|eng d
035  ‎$a‎6800013446
040  ‎$a‎MAP‎$b‎spa
084  ‎$a‎872
24510‎$a‎Biological safety cabinets‎$b‎: decontamination or sterilization with paraformaldehyde‎$c‎Richard Fink... [et al.]
520  ‎$a‎Bacillus subtilis var. niger spores were used to determine the exposure time for formaldehyde decontamination of biological safety cabinets. Formaldehyde contact times less than 3 hr were insufficient for sterilization. A contact time of 4 hr or more resulted in a reproducible killing of the spore strips placed inside the cabinets. At 6 hr sufficient formaldehyde had diffused through the high efficiency particulate air (HEPA) filter to sterilize the strips with lower spore counts. A minimum of 5 to 6 logs of killing were observed after 4 to 6 hr of treatment
65011‎$0‎MAPA20080585679‎$a‎Higiene industrial
65011‎$0‎MAPA20080605278‎$a‎Contaminantes químicos
65011‎$0‎MAPA20080613617‎$a‎Contaminantes biológicos
65011‎$0‎MAPA20080562595‎$a‎Formaldehido
65011‎$0‎MAPA20080568887‎$a‎Esterilización
65011‎$0‎MAPA20080582241‎$a‎Recintos cerrados
7001 ‎$0‎MAPA20080068660‎$a‎Fink, Richard
7400 ‎$a‎American Industrial Hygiene Association journal
7730 ‎$t‎American Industrial Hygiene Association journal‎$d‎Akron, Ohio‎$g‎Vol. 49, nº 6, June 1988 ; p. 277-279