Pesquisa de referências

Gravimetric determination of airborne dust by using a filter cartridge inside a closed-face, 37 mm polystyrene cassette

<?xml version="1.0" encoding="UTF-8" standalone="no"?>
<rdf:RDF xmlns:rdf="http://www.w3.org/1999/02/22-rdf-syntax-ns#" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance">
<rdf:Description>
<dc:creator>Puskar, Mark A.</dc:creator>
<dc:date>1992-11-01</dc:date>
<dc:description xml:lang="es">Sumario: This paper reports on a new filter cartridge device that was developed and evaluated to estimate the concentration of airborne total dusts over a larger linear range while eliminating negative bias errors associated with both sample losses on the inside of the cassette's top during sampling and shipment and with sample losses during filter removal and handling</dc:description>
<dc:identifier>https://documentacion.fundacionmapfre.org/documentacion/publico/es/bib/47997.do</dc:identifier>
<dc:language>eng</dc:language>
<dc:rights xml:lang="es">InC - http://rightsstatements.org/vocab/InC/1.0/</dc:rights>
<dc:subject xml:lang="es">Higiene industrial</dc:subject>
<dc:subject xml:lang="es">Contaminación atmosférica</dc:subject>
<dc:subject xml:lang="es">Control de la contaminación</dc:subject>
<dc:subject xml:lang="es">Polvo</dc:subject>
<dc:subject xml:lang="es">Filtros</dc:subject>
<dc:subject xml:lang="es">Muestreos</dc:subject>
<dc:subject xml:lang="es">Muestreador</dc:subject>
<dc:type xml:lang="es">Artículos y capítulos</dc:type>
<dc:title xml:lang="es">Gravimetric determination of airborne dust by using a filter cartridge inside a closed-face, 37 mm polystyrene cassette</dc:title>
<dc:title xml:lang="es">Título: American Industrial Hygiene Association journal</dc:title>
<dc:relation xml:lang="es">En: American Industrial Hygiene Association journal. - Akron, Ohio. - Vol. 53, nº 11, November 1992 ; p. 692-698</dc:relation>
</rdf:Description>