An Improved instrument for the in vivo detection of lead in bone

Fichero PDF / PDF file
MARC record
Tag12Value
LDR  00000nab a2200000 i 4500
001  MAP20071015284
003  MAP
005  20080418114139.0
007  hzruuu---uuuu
008  931104e19930701gbr|||| | |00010|eng d
040  ‎$a‎MAP‎$b‎spa
084  ‎$a‎931
1001 ‎$0‎MAPA20080044268‎$a‎Gordon, C.L.
24513‎$a‎An Improved instrument for the in vivo detection of lead in bone‎$c‎C.L. Gordon, D.R. Chettle, C.E. Webber
520  ‎$a‎An improved instrument for the fluorescence excitation measurement of concentrations of lead in bone has been developed. This is based on a large area high purity germanium detector and a point source of cadmium. The source is positioned in a tungsten shield at the centre of the detector face such that 88keV photons cannot enter the detector directly. In vivo measurements are calibrated with plaster of Paris phantoms. This will allow studies of the natural history of non-occupational lead accumulation in normal subjects and should permit investigations in subjects poisoned with lead
65011‎$0‎MAPA20080548124‎$a‎Medicina
65011‎$0‎MAPA20080586546‎$a‎Nuevas tecnologías
65011‎$0‎MAPA20080560782‎$a‎Terapéutica
65011‎$0‎MAPA20080567576‎$a‎Traumatología
7001 ‎$0‎MAPA20080065188‎$a‎Chettle, D.R.
7001 ‎$0‎MAPA20080058937‎$a‎Webber, C.E.
7400 ‎$a‎British journal of industrial medicine
7730 ‎$t‎British journal of industrial medicine‎$d‎London and Margate‎$g‎nº 7, July 1993 ; p. 637-641