Seção: ArtigosTítulo: Performance characteristics of control devices to maintain constant face velocity for laboratory exhaust hoods / Richard P. Garrison... [et al.]Notas: Sumario: Performance characteristics were investigated for six commercial devices designed to maintain constant face velocity (FV) with changing sash position for a laboratory exhaust hood. Testing involved displacement of the hood sash upward and downward and monitoring hood and controller performance. Some FV deviations were too great to assure adequate control for potential exposures to highly toxic materialsRegistros relacionados: En: American Industrial Hygiene Association journal. - Akron, Ohio. - Vol. 50, nº 9, September 1989 ; p. 501-504Materia / lugar / evento: Higiene industrialContaminantes químicosSustancias tóxicas y peligrosasControl de la contaminaciónVentilaciónVelocidadAireLaboratoriosOtros autores: Garrison, Richard P. Títulos secundários: Título: American Industrial Hygiene Association journal Outras classificações: 872Direitos: In Copyright (InC)